Figure 1: The functional sensitivity of Elecsys Anti-TSHR at approx. 0.9 IU/L is significantly below the cut-off (≥1.75 IU/L), allowing clear differentiation of pathological results.
Elecsys Anti-TSHR (TRAK) is a fully automated test for detection of autoantibodies to the TSH receptor
Elecsys Anti-TSHR (TRAK) is a fully automated test for detection of autoantibodies to the TSH receptor
Figure 1: The functional sensitivity of Elecsys Anti-TSHR at approx. 0.9 IU/L is significantly below the cut-off (≥1.75 IU/L), allowing clear differentiation of pathological results.
Figure 2: The high sensitivity of 97% and high specificity of 99.5% of the Elecsys Anti-TSHR assay support clinical accuracy
The clinical study comprised:
Using a cutoff of 1.75 IU/L a clinical sensitivity of 97% and a specificity of 99% was obtained.
* LoB = Limit of Blank; LoD = Limit of Detection; LoQ = Limit of Quantitation with a total allowable error of ≤20 %.