For in vitro diagnostic use. Instrument cobas e 801 IVD cobas® e 801 analytical unit INS_2202 07682913001 cobas e 801 module cobas e 801 analytical unit 04015630946198 Instruments Not Available Not Available 08454345001 cobas e 801 analytical unit cobas e 801 analytical unit 07613336226636 Instruments Not Available Not Available The e 801 module is a fully automated, high throughput immunology analyzer for a large array of quantitative and qualitative in vitro tests including cobas e flow tests. en Specifications of the consumables system Capacity of magazine lifter15 magazinesAssayCups per magazine105AssayTips per magazine105Max. number of AssayCups in magazine lifter1575Max. number of AssayTips in magazine lifter1575Max. walkaway time5 h 15 minNumber of solid waste containers2Capacity of each solid waste container15 magazinesCapacity of magazine waste compartment15 magazinesSpecifications of the reagent system Reagent cassette typecobas e pack greenReagent loading/unloadingAutomaticCap pre-openerYesReagent identificationRFIDCapacity of reagent disk48 cassettesReagent storage temperature5–10°C (41–50°F)Reagent pipetting volume6–60 μLMicrobead pipetting volume6–60 μLReagent volume controlSoftware-controlled test countEvaporation protectionYesSpecifications of the sampling system Sample typesSerum/plasmaUrineCerebrospinal fluidSupernatantOthersWhole bloodOral fluidHemolysateAmniotic fluidProcessed stoolSample pipetting principleOne disposableAssayTip/pipettingSampling cycle time12 s per module24 s per measuring channelSample pipetting volume4–60 μLLiquid level detectionYesClot detectionYesFoam detection with cameraYesSpecifications of the reaction system Number of applications that can be loaded on the instrument200Number of incubator disk positions94Reaction volume120 μLIncubator temperature37°C ± 0.3°C (98.6°F ± 0.5°F)Reaction times for tests9/18/27 minAutomatic dilution1-step, 2-step, 3-stepDilution ratio1:1 to 1:27000MixerVortexSpecifications of the ECL measuring system Measuring cellECL measuring cell V7.0Number of channels2Cycle time per channel24 sOverall cycle time (2 channels)12 sMax. throughput300 tests/hTemperature of detection unit28°C ± 0.5°C (82.4°F ± 0.9°F)Temperature control (detection unit)Aluminum panel, air circulationTests per PreClean bottle≥ 2000Tests per ProCell bottle≥ 1000Tests per CleanCell bottle≥ 800 en