BenchMark Special Stains

IVD For in vitro diagnostic use.

The BenchMark Special Stains system enhances workflow with proven technology

The BenchMark Special Stains system offers advanced automation for histology special stains. By fully automating the process from baking to staining, the BenchMark Special Stains system enables the consistent and efficient quantitative detection of targets, aiding accurate diagnosis.

Productivity features such as the ability to run any test next to any other test and waste fluid sensing optimize workflow. Reducing manual processes and improving lab capabilities allows your skilled laboratory professionals to focus on higher-value contributions.

Additionally, an intuitive user interface (Ventana System Software) connects up to eight BenchMark Special Stains and IHC/ISH instruments, enabling seamless integration and easy adoption for lab personnel.

A lab technologist holds up a case of tissue slides that were stained with the BenchMark Special Stains instrument.
Benefits at a glance

Benefits at a glance

Overview

Ordering information

Product specification Value
General characteristics
Fully automated Baking, deparaffinization and staining of special stains
Slide carousel 1–20 slides with independent temperature control for each position
Reagent carousel 25 reagent positions
Slides 25 mm x 75 mm or 26 mm x 76 mm Superfrost Plus
Bulk fluids Up to four bulk fluids in 3–6 L onboard containers
DI water quality specifications CLSI guideline C3-A4 Clinical Laboratory Reagent Water (CLRW)
Modularity 1–8 BenchMark instruments may be controlled from one PC
Configuration Freestanding or modified benchtop installation
Certifications CSA, EMC, CE compliant
Ordering code 06657389001
Temperature control
Instrument operational temperature 37°C–75°C (98°F–167°F)
Environmental requirements
Heat 1200 BTU/Hr
Operating temperature 15°C to 30°C (64°F to 90°F)
Humidity 20% to 80%, non-condensing
Electrical specifications
Voltage 100 VAC +/- 10% (Japan), 120 VAC +/- 10% (US & Canada), 230 VAC +/- 10% (Europe)
Current 4 Amps
Frequency 50/60 Hz
Power connection Standard 3-prong grounded
Fuse Waste module: 6.3 Amps, 250 Volts, 5 mm x 20 mm, fast-acting
Instrument physical characteristics
Size (W x D x H) 50.8 cm x 63 cm x 135.4 cm (20 in x 25 in x 53.3 in)
Weight 76.7 kg (165.7 lbs)
Clearances Rear: 15.2 cm (6 in)
Top: 30.5 cm (12 in)
Left side: 7.6 cm (3 in)
Right side: 61 cm (24 in) to open door to 90°
Right side: 101.6 cm (40 in) to open door to 180°
Front: 108 cm (42.5 in)