Built upon the proven technology of the BenchMark ULTRA system, BenchMark ULTRA PLUS delivers an enhanced lab experience and an optimized workflow with flexible solutions. This easy-to-use solution helps histotechnologists and anatomic pathology laboratories manage activities efficiently and confidently, saving on personnel time and resource costs.

 

Key advantages of the BenchMark ULTRA PLUS include:

  • Workflow efficiency through remote connectivity
  • Fewer instrument touchpoints
  • Software interface usability
  • On-instrument run management capabilities
  • Electronic inventory management
  • Quality assurance
  • Useful reporting metrics
  • Over 200 assays in key diseases areas, including high medical value assays
Tissue image
Learn more about the BenchMark ULTRA PLUS features

Optimized workflow

Improve usability while enabling time for more value-added activities through:

  • New light notifications communicate instrument status from across the lab
  • New retractable work surface gives technologists additional space to work while at the instrument
  • New bulk reagent color-coded clips make bulk reagent loading easier and helps minimize errors
  • New remote access allows you to view, manage, complete, and print system data remotely
  • New integrated communication allows streamlined data management and inventory tracking across all host computers
BenchMark ULTRA PLUS reagent colored clips

Flexible solutions

New features based on user feedback:

  • New configuration has a decreased height compared to the BenchMark ULTRA
  • New integrated touchscreen reduces bottlenecks by allowing many functions to be done at the instrument
  • VSS 14.x software allows for easier navigation of inventory management and ordering through a new filter as well as lot-to-lot validation on the same instrument
  • New proprietary waste management* solution degrades DAB (3,3'-diaminobenzidine) to less than 5 ppb meeting the most stringent municipality requirements for nonhazardous waste

*It is the responsibility of each customer to ensure compliance with all applicable laws and regulations related to the disposal of waste.

BenchMark Ultra Plus touchscreen close up

Quality is at the core

Proven BenchMark technologies deliver the right stain at the right time:

  • Flexible, independent temperature control for individual slide staining optimization
  • Proven staining technology of air vortex mixing with LCS create a reaction chamber allowing full slide coverage with 100ul of reagent
  • New slide drawer design mitigates fluid ingress and egress resulting in continuous and highly-reliable drawer performance
  • VSS 14.x software has new and expanded reporting capabilities including quality assurance/quality control (QA/QC)
BenchMark ULTRA PLUS glass stain

Would you like to know more about BenchMark ULTRA PLUS?

 

Submit your details in order to be contacted by a Roche representative with more details about BenchMark ULTRA PLUS and its availability in your region.

Form Successfully Submitted!
Thank you for your submission!
text

Overview

Ordering Information

General characteristics

  • Fully automated

    Baking, deparaffinization, cell conditioning and staining, including IHC, ISH, SISH, immunofluorescence, multiparameter IHC staining (dual, triple), counterstain and titration

  • Slide position

    1–30 slides with independent processing/functionality and temperature control for each position

  • Reagent carousel

    35 reagent positions

  • Slides

    25 x 75 mm, 1 x 3 mm or 26 x 76 mm positively charged slides

  • Bulk solutions

    Up to 7 different bulk solutions can be changed without process interruption

    Only required bulk reagents needed on board

  • Waste

    Two 20 L carboys for processed aqueous waste

    Two 5 L bottles for separated LCS waste

    Virtually undetectable DAB in waste: <5 parts per billion

  • DI water supply

    NCCLS Type II water or equivalent (deionized water)

  • Temperature control

    Ambient to 100°C

  • Modularity

    1 to 4 BenchMark ULTRA PLUS instruments may be controlled from one BenchMark ULTRA PLUS computer system

  • Configuration

    Free standing

  • Certifications

    EN 61010-1, IEC/EN 61010-2-010, IEC/EN 61010-2-101, EN 61326-2-6

  • Catalog number

    N750-BMKU-FS 05342716001

Physical characteristics

  • Size (W x D x H)

    44 X 35.25 X 59.15 in (111 X 89.5 X 150.2 cm)

  • Weight

    Instrument with reagents: 854.9 lbs (387.8 kg) full fluids, bulk, reservoirs and UV. Instrument dry: 606 lbs (275 kg)

  • Clearances

    Top 16.6 in (42.2 cm)
    Sides 4 in (10.2 cm)
    Rear 6 in (15.2 cm)

Environmental requirements

  • BTU/Hour output

    400 BTU/Hr idle, 1000 BTU/Hr running

  • Room temperature

    68°F to 90°F (20°C to 32°C)

  • Humidity

    20% - 88%, non-condensing

Electrical specifications

  • Voltage

    ~100–240V

  • Current

    5-10 Amps

  • Frequency

    50/60 Hz