Ordering Information

Technical Documents

Specifications of the reagent system

  • Reagent pack types

    cobas e pack green

  • Reagent loading / unloading

    Manual

  • Reagent Identification

    RFID

  • Capacity of reagent disk

    28 reagent packs

  • Reagent storage temperature

    5 – 10 °C (41 – 50 °F)

Specifications of the sampling system

  • Sampling cycle time

    30 seconds

  • Sample pipetting volume

    4 – 60 μL (1 μL steps)

  • Sample Liquid level detection

    Available

  • Sample clot detection

    Available

  • Sample air aspiration detection

    Available

Specifications of the reaction system

  • Number of incubator disk positions

    38

  • Reaction volume

    120 μL

  • Incubator temperature

    37 °C ± 0.3 °C (98.6 °F ± 0.5 °F)

  • Reaction times for tests

    9/18/27 min

  • Mixer

    Vortex

Specifications of the ECL measuring system

  • Measuring Cell

    ECL measuring cell

  • Number of measuring cells

    1

  • Maximum throughput*

    120 tests/hour*

* Throughput may differ based on the mix of test orders per sample