{ "ProductData": { "ID": "INS_2009", "ProductType": "Instrument", "BrandName": "cobas u 411 urine analyzer", "ProductNameAddition": "", "RegulatoryDisclaimer1": "For in vitro diagnostic use.", "DisclaimerGroup1": "IVD", "RegulatoryDisclaimer2": null, "DisclaimerGroup2": null, "RegulatoryDisclaimer3": null, "SampleType": [ ], "LicenseDisclaimers": [ ], "RelatedLinks": "", "Clone": "", "ControlTissue": [ "" ], "ISOtypes": "", "Species": [ "" ], "StainLocalization": [ "" ], "ProductNameGlobal": "cobas u 411" }, "ProductImageDetails": { "ImagePath": "https://pim-media.roche.com/Images/INS_2009_cobas_u_411.jpg", "ImageType": "Image main" }, "Product2Taxonomy": { "Product2TaxonomyReferences": [ { "StructureSystemIdentifier": "1", "StructureSystemName": "GPCH", "NodeID": "2159", "StructureNodeStatus": "", "NodeName": "" }, { "StructureSystemIdentifier": "OWP_Family", "StructureSystemName": "Product Families", "NodeID": "110", "StructureNodeStatus": "valid", "NodeName": "cobas" }, { "StructureSystemIdentifier": "OWP_Product_Types", "StructureSystemName": "Product Types", "NodeID": "10-000-00", "StructureNodeStatus": "valid", "NodeName": "Analyzer Instruments and Systems" }, { "StructureSystemIdentifier": "Lab_Type", "StructureSystemName": "Lab Types", "NodeID": "050-00", "StructureNodeStatus": "valid", "NodeName": "Core Lab" }, { "StructureSystemIdentifier": "Product_Solutions", "StructureSystemName": "Product Solutions", "NodeID": "120", "StructureNodeStatus": "valid", "NodeName": "Urinalysis" }, { "StructureSystemIdentifier": "Applications", "StructureSystemName": "Applications", "NodeID": "99-00-00", "StructureNodeStatus": "invalid", "NodeName": "" }, { "StructureSystemIdentifier": "OWP_Techniques", "StructureSystemName": "Techniques", "NodeID": "999-00", "StructureNodeStatus": "invalid", "NodeName": "" }, { "StructureSystemIdentifier": "Disease_Areas", "StructureSystemName": "Disease Areas", "NodeID": "99-00-00", "StructureNodeStatus": "invalid", "NodeName": "" }, { "StructureSystemIdentifier": "Pathogens", "StructureSystemName": "Pathogens", "NodeID": "99-00-00", "StructureNodeStatus": "invalid", "NodeName": "" }, { "StructureSystemIdentifier": "Health_Topics", "StructureSystemName": "Health Topics", "NodeID": "99-00-00", "StructureNodeStatus": "invalid", "NodeName": "" } ] }, "Product2Materials": { "P2MaterialReferences": [ { "MaterialNum": "04906969001", "MaterialDescription": "cobas u 411 urine analyzer", "RegisteredProductName": "cobas u 411 urine analyzer", "GTIN": "04015630925636", "ProductCategoryText": "Instruments", "OldMaterialNumber": "", "PackSizePIM360": "Not Available", "PackSizeDescPIM360": "Not Available", "MaterialAnnotation": "", "ReadyForUse": "", "OrderInformation": "" } ] }, "Product2Products": { "Product2ProductReference": [ ] }, "ProductSpec": [ { "ProductSpecVariant": { "Chapters": [ { "Name": "Intended Use", "Value": "El analizador cobas u 411 es un sistema de análisis de orina semiautomático destinado a la determinación cualitativa o semicuantitativa in vitro de analitos de orina, incluidos la densidad (SG), el pH, los leucocitos, los nitritos, las proteínas, la glucosa, las cetonas, el urobilinógeno, labilirrubina y los eritrocitos. El analizador también determina el color de la orina.

El analizador cobas u 411 utiliza las tiras reactivas Combur10 Test M (consulte Marcas comerciales en la página 3).

El analizador cobas u 411 ha sido concebido para un uso óptimo con cargas de trabajo de entre 30 y 100 muestras de orina al día. Únicamente personal formado que trabaje en un laboratorio profesional puede utilizar el analizador cobas u 411. El analizador ha sido desarrollado y diseñado para analizar muestras de orina solamente.

Es importante que los usuarios lean atentamente este manual antes de utilizar el analizador.

Cualquier descuido al seguir las instrucciones del Manual del operador puede ocasionar riesgos para su seguridad.", "Language": "es", "Country": "XG", "Code": "Intended Use" }, { "Name": "Specification", "Value": "Analyzer specifications
 
System descriptionContinuous load urine analyzer for test stripsParameters:
  • Spec. gravity
  • pH
  • Leukocytes
  • Nitrite
  • Protein
  • Glucose
  • Ketone
  • Urobilinogen
  • Bilirubin
  • Erythrocytes
  • Color
PrecisionIntra-System Precision
  • Control Test M
Inter-System Precision
  • Control Test M

≤ 0.5% Remission

≤ 1.5% Remission
Robustnessaccepted bending of test strips
-wet
-dry 2 mm

4 mm
2 mm
Software and data handling
CPUs


 
Software version 3.2:
  • Intel XScale PXA255
  • Cygnal 8051
Software version 3.3:
  • Intel XScale PXA168
  • Cygnal 8051
Operating systemLINUX
MemorySoftware version 3.2:
  • 32 MB Flash-Memory
  • 32 MB DRAM
Software version 3.3:
  • 128 MB Flash-Memory
  • 128 MB DRAM
Bus architectureCAN fieldbus
Data storage
  • internal
  • external

Flash-Memory
USB port
InterfacesUSB Port: 1 x USB
RS232 ports: 3 x RS232 (Sediment Terminal, Barcode Scanner, Host)
DisplaySoftware version 3.2: 5.7 inch (¼ VGA - 320 x 240 pixels) passive matrix touch screen, blue/white. Software version 3.3: 5.7 inch (¼ VGA - 320 x 240 pixels) passive matrix touch screen color.
PrinterInternal thermal-printer (112 mm)
Samples
Sample handlingManual by user
Throughput (theoretical throughput)600 test strips per hour
Incubation time60 seconds
Instrument cycle time6.0 seconds
CalibrationReference Strip on board for internal calibration

Recommended calibration with calibration strip: once a month
Control Test M


Control Test M
Measuring unitLight source
Wavelength
Sensor
20 LEDs
470 nm, 555 nm, 620 nm
11 wide range photo sensors
Physical dimensionsHeight
Width
Depth
Weight 
260-360 mm
425 mm 
340-375 mm
approx. 12 kg
Power requirementsLine voltage
Line frequency
Power Consumption
Overvoltage Category
Fuse
100 V – 125 V (-15% / +10%) and 200 V – 240 V (-15% / +10%)
50 Hz (+/- 5%) or 60 Hz (+/-5%)
60 VA
II (IEC 61010-1)
T 2.5A 250V
 
Environmental conditionsTemperature
  • Running conditions
  • Transportation and storage
Humidity
  • Running conditions
  • Transporation and storage
Pollution degree
Altitude
 
  • 15 – 32°C (59 – 89.6°F)
  • -25 – 60 °C (-13 – 140 °F)
10% – 95% at no condensation
  • 30% - 80%, non-condensing
  • 10% - 95%, non-condensing
2 (IEC 61010-1)
0 – 2000 m, air pressure 80 to 106 kPa
", "Language": "en", "Country": "XG", "Code": "Specification" } ] } } ] }

Analizador de orina cobas u 411

Gestión de resultados consolidados

IVD For in vitro diagnostic use.
Analizador de orina cobas u 411
¿Tienes alguna pregunta?
Ponte en contacto con nosotros

Procesamiento rápido y eficaz

 

Solo 6 segundos para una lectura de resultados: Esta solución ofrece una carga continua de tiras reactivas sin esperar al ciclo de medición. La solución también ofrece una medición de resultados estandarizados mediante la eliminación del factor humano.

 

Resultados consolidados

 

El terminal de sedimentos permite la consolidación de datos de las tiras reactivas y de microscopía en un resultado común.

 

Notificación flexible y exhaustiva

 

Creación de informes de resultados a través de una impresora integrada para imprimir o transmitir datos en tiempo real a un LIS.

analizador de orina cobas u 411 en el laboratorio

Características del producto1

 

  • Cargas de trabajo: 30 - 100 muestras al día
  • Carga continua de tiras reactivas
  • El tiempo de ciclo de 6 segundos permite procesar un gran número de muestras y un procesamiento rápido
  • Lector de código de barras opcional para simplificar los pasos manuales
  • El uso de la prueba Combur10 M con tecnología de sellado de red proporciona las ventajas bien conocidas de las tiras reactivas de Roche

 

Datos consolidados, mejora del flujo de trabajo del laboratorio

 

Los resultados de microscopía manual pueden introducirse a través de la terminal de sedimentos y transferirse al analizador cobas u 411. Esto permite la documentación sencilla y una mejor visión general de los registros de los pacientes, con impresión única de la información de las tiras reactivas y microscopía.

explicación del flujo de trabajo del analizador de orina cobas u 411

Referencias bibliográficas

 

  1. cobas u 411 system, O. M., 2014. eLabDoc. [Online] Disponible en: https://dialog1.roche.com.

Resumen

Pedidos

Documentos técnicos

error errorMessage
Se ha producido un problema al cargar esta sección.
Vuelve a cargar esta página o inténtalo de nuevo más tarde.