Your USB flash drive must meet these requirements:
Parameter | Value |
---|---|
USB Formating* | FAT32 |
USB Allocation Size | 4GB minimum |
USB Version | USB 2.0 compatible minimum |
*The USB should be formatted prior to copying the update package onto it. Please confirm your USB is empty prior to formatting. To format your USB flash drive:
|
Update script: Cobas Liat Strep_A Script v1.29.zip
Note: The update package must be in the root directory of the USB flash drive (exactly as shown in the image below) in order for the software to install correctly on an analyzer.
Log into your LIAT as an ADMIN user.
From the Main screen, insert your script USB flash drive at the rear of the analyzer.
A message dialog box should automatically open after inserting your script USB. Follow the on screen prompts for assay script installation.
Choose Confirm, and enter your password.
Choose Yes to confirm that you want to install the assay script. A progress screen is displayed.
Follow the dialog box: Remove the USB flash drive. Wait 5 seconds. Then choose Confirm to continue. The script installation completes.
When a successful script installation is displayed in the Status column, choose the Confirm button.
After installing the new script, activation is required by the user. The new script must be registered through activation within 30 days. You can use the unregistered assay during the 30-day registration period. But, once this period has expired, the assay can no longer be used for testing until the activation has been completed.
Note: If you are unable to select Activate, confirm that you are logged in as an ADMIN user.
If this is the first time SASA has been installed on your Liat, or if at any time you are adding new users, you will need to update your user accounts in order to run the assay.
Note: You are able to transfer user accounts from one Liat to another using a USB flash drive. (Remember the SASA script USB flash drive you generated should be used for no other purpose than installing the SASA script.)
To update your user accounts to allow SASA for use:
Login as an Admin User.
Select Settings.
Select User Management.
Select the desired user.
Scroll down to Assays and choose Edit.
On the Assay Permissions screen, toggle SASA to Allowed and select Apply.
To transfer user accounts from one Liat to another:
First export the desired user list
Insert your USB flash drive.
Select Tools.
Select Export Data.
Scroll down to Export Users and choose Select.
Wait until a message is displayed that the users were exported successfully and then choose Confirm.
Then import the user list to any newly updated Liat
Login as an Admin User.
Insert your USB flash drive.
Select Tools.
Select Import Data.
Scroll down to Import Users and choose Select. Enter your password and choose Enter.