Elecsys® HIV Duo

IVD For in vitro diagnostic use.
Elecsys<sup>®</sup> HIV Duo
Immunoassay for the qualitative determination of HIV p24 antigen and antibodies to HIV

The human immunodeficiency virus (HIV) is the causative agent of acquired immunodeficiency syndrome (AIDS) and has been a major global burden for over three decades.1,2 HIV is transmitted through sexual contact, contaminated blood and blood products or from an HIV-infected mother to her child before, during and after birth.3 Diagnosis of an HIV infection can be made as early as 2 - 3 weeks after infection, based on the detection of HIV p24 antigen in the blood.4,5 Anti-HIV antibodies are detectable in serum from around 4 weeks post-infection.4,6 

With the Elecsys® HIV Duo assay, HIV-1 p24 antigen (HIV Ag), as well as antibodies to HIV-1 and HIV-2 (anti-HIV) can be detected in parallel with two separate determinations. On the basis of these determinations, the Elecsys® HIV Duo main result is calculated automatically by the analyzer. The subresults for HIV Ag and anti-HIV can be used as an aid in the selection of the confirmation algorithm for reactive samples.7

Elecsys® HIV Duo

  • Systems

    cobas e 801 module

  • Testing Time

    18 minutes

  • Test principle

    Double antibody or antigen sandwich immunoassay for the detection of HIV antigen and anti-HIV antibodies, respectively.

  • Calibration

    Individual 2-point calibration for HIV antigen and anti-HIV antibodies

  • Interpretation

    <1.0 = non-reactive 
    ≥1.0 = reactive

  • Sample material

    Serum collected using standard sampling tubes or tubes containing separating gel. Li-heparin, Na-heparin, K2-EDTA, K3-EDTA, ACD, CPD, CP2D, CPDA, potassium oxalate and Na-citrate plasma as well as Li-heparin has plasma tubes containing separating gel

  • Sample volume

    60 μL (2 x 30 μL)

  • Onboard stability

    16 weeks

  • Intermediate precision in positive samples

    HIV Ag- 1.5-2.2 CV%
    Anti-HIV- 1.2-1.8 CV%

  • Clinical sensitivity

    100 % (total n = 1549, HIV‑1 confirmed positive pediatric subjects, n=51, HIV‑1 confirmed positive pregnant women subjects, n=59, HIV‑1 group M subtype specimens, n=90, HIV‑1 group O specimens , n=50, and HIV‑1 p24 antigen

  • Clinical specificity

    99.90% (n = 6894 low risk population)

  • Analytical sensitivity

    0.392 IU/mL

References

  1. Maartens, G., Celum, C., Lewin, S.R. (2014). HIV infection: epidemiology, pathogenesis, treatment, and prevention. Lancet 384, 258-71.
  2. Killian, M.S., Levy, J.A. (2011). HIV/AIDS: 30 years of progress and future challenges. Eur J Immunol 41, 3401-11.
  3. Shaw, G.M., Hunter, E. (2012). HIV transmission. Cold Spring Harb Perspect Med. 2:a006965.
  4. Fiebig, E.W., Wright, D.J., Rawal, B.D. et al. (2003). Dynamics of HIV viremia and antibody seroconversion in plasma donors: implications for diagnosis and staging of primary HIV infection. AIDS 17, 1871-9.
  5. Busch, M.P., Lee, L.L., Satten, G.A. et al. (1995). Time course of detection of viral and serologic markers preceding human immunodeficiency virus type 1 seroconversion: implications for screening of blood and tissue donors. Transfusion 35, 91-7.
  6. Guertler, L., Muehlbacher, A., Michl, U. et al. (1998). Reduction of the diagnostic window with a new combined p24 antigen and human immunodeficiency virus antibody screening assay. Journal of Virological Methods 75, 27-38.
  7. HIV Duo Package Insert- 2020-05 V1.0.

Would you like to know more about Elecsys® HIV Duo?

Please submit your information in the following form to be contacted by a Roche representative with more details about Elecsys® HIV Duo.

Form Successfully Submitted!
Thank you for your submission!
text

Overview

Detailed Specifications

Ordering Information

Compatible Instruments

...
    ...

    Technical Documents

    error errorMessage
    Sorry, we couldn't find the content you are looking for
    Please try again later