The cobas b 221 blood gas system sets
the gold standard for pleural fluid pH testing
 

Measuring pleural fluid pH is a clinically useful tool for diagnosing and managing
patients with pleural effusions and can be especially important in critical care
situations, when dependable performance is essential.

 

Do you have any questions?
Contact us

 

Measured parameters

Blood gas module Total specified range
pH 6.0 - 8.0
PCO2 4.0 - 200 mmHg
PO2 0 - 800 mmHg
Electrolyte module
Na+ 20.0 - 250 mmol/L
K+ 0.2 - 20 mmol/L
Ca2+ 0.1 - 4.0 mmol/L
Cl- 20.0 - 250 mmol/L
Hct 10% - 80%
Hemoglobin module
tHb 3 - 25 g/dL
SO2 50% - 100%
Metabolite module
Glu 0.5 - 40 mmol/L
Lactic acid 0.2 - 20 mmol/L
Urea (BUN) 0.5 - 30 mmol/L
CO-oximetry module
tHb-COOX 3 - 25 g/dL
O2Hb 0% - 100%
HHb 0% - 100%
COHb 0% - 100%
MetHb 0% - 100%
SO2 COOX 0% - 100%
Bilirubin (neonatal) 3 - 50 mg/dL
Barometric pressure (Baro) 450 - 800 mmHg

 

General characteristics

Calculated parameters

H+, cHCO3-, ctCO2(P), FO2Hb, BE, BEecf, BB, SO2, P50, ctO2, ctCO2(B), pHst, cHCO3-st, PAO2, AaDO2, a/AO2, avDO2, RI, shunt, nCa2+, AG, pHt, H+t, PCO2t, PO2t, PO2t, AaDO2t, a/AO2t, RIt, Hct(c), MCHC, BO2, BEact, osmolality, OER, heart minute volume (Qt), P/F index

Sample types

Whole blood, serum, plasma, dialysate, QC material

Data processing

Industrial standard PC

Monitor

Built-in flat color TFT-LCD 10.4-inch screen (touchscreen)

Thermal printer

Built-in 111 mm, graphical capability

Bar code scanner

Standard accessory

Electrical requirements

Power rating

100 - 240V (+6% - 10% permission to tolerance), 200W, 50/60 Hz autoselecting

Ambient temperature

+15 to +31 C (59 to 89.6 F)

Relative humidity, not condensed

15 < T < 31 C: 20% - 85%

Options

Built-in AQC

Automatic quality control system, with room for up to 120 QC ampoules

Test certificate

FDA

510 (k)

UL

UL3101-1

CE-conformity

IVD-Directive 98/79/EC (IEC 1010-1 / EN 61010-1 / EN 61010-2-101)

Dimension/weight

Width

< 51 cm

Height

< 59 cm (in operating situation)

Depth

< 60 cm (with power supply)

Weight (without solutions, without AutoQC)

45 kg

Calibration

Intervals, Duration

System calibration

Every 24 hours (programmable 8, 12, or 24 hours)

1-point calibration

Every 60 minutes (programmable 30 to 60 minutes)

2-point calibration

Every 12 hours (programmable 4, 8, or 12 hours)

Warm-up

Power on with calibration < 43 minutes 
Power fail < 1 minutes < 2.5 minutes

 

Maintenance

Maintenance frequency Time required Required maintenance
Daily 1 minute
  • Check fill levels
  • Check printer paper
Weekly 3 - 5 minutes
  • Clean fill port and sample drip tray
  • Clean screen
Quarterly 5 - 10 minutes
  • Clean T&D disk
  • Perform COOX calibration
Biannually 1 minute
  • Change peristaltic pump tubing
  • Change air filter
As needed  
  • Change bottle or pack
  • Change printer paper
  • Clean modules and sample paths
  • Change fill port
  • Clean bottle compartment
  • Decontaminate surfaces
  • Change AutoQC mats